Comparison of imaging parameters between OCT, GDx and HRT in the Northern Finland birth cohort eye study

Stoor, Katri; Karvonen, Elina; Leiviskä, Ilmari; Liinamaa, Johanna; Saarela, Ville (2021-10-19)

 
Avaa tiedosto
nbnfi-fe2023031431537.pdf (1.223Mt)
nbnfi-fe2023031431537_meta.xml (44.78Kt)
nbnfi-fe2023031431537_solr.xml (42.37Kt)
Lataukset: 

URL:
https://doi.org/10.1111/aos.15046

Stoor, Katri
Karvonen, Elina
Leiviskä, Ilmari
Liinamaa, Johanna
Saarela, Ville
John Wiley & Sons
19.10.2021

Stoor, K., Karvonen, E., Leiviskä, I., Liinamaa, J. and Saarela, V. (2022), Comparison of imaging parameters between OCT, GDx and HRT in the Northern Finland birth cohort eye study. Acta Ophthalmol, 100: e1103-e1111. https://doi.org/10.1111/aos.15046

https://rightsstatements.org/vocab/InC/1.0/
© 2021 Acta Ophthalmologica Scandinavica Foundation. Published by John Wiley & Sons Ltd. This is the peer reviewed version of the following article: Stoor, K., Karvonen, E., Leiviskä, I., Liinamaa, J. and Saarela, V. (2022), Comparison of imaging parameters between OCT, GDx and HRT in the Northern Finland birth cohort eye study. Acta Ophthalmol, 100: e1103-e1111, which has been published in final form at https://doi.org/10.1111/aos.15046. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Use of Self-Archived Versions. This article may not be enhanced, enriched or otherwise transformed into a derivative work, without express permission from Wiley or by statutory rights under applicable legislation. Copyright notices must not be removed, obscured or modified. The article must be linked to Wiley’s version of record on Wiley Online Library and any embedding, framing or otherwise making available the article or pages thereof by third parties from platforms, services and websites other than Wiley Online Library must be prohibited.
https://rightsstatements.org/vocab/InC/1.0/
doi:https://doi.org/10.1111/aos.15046
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe2023031431537
Tiivistelmä

Abstract

Purpose: The purpose of the study was to assess and compare the optic nerve head (ONH) and retinal nerve fibre layer (RNFL) parameters and image quality parameters obtained by Cirrus HD-OCT, GDxECC and HRT3 in a population-based screening study.

Methods: This analysis examined 2566 subjects taking part in the Northern Finland Birth Cohort Eye study. Images with spectral domain OCT (Cirrus HD-OCT 4000), scanning laser ophthalmoscopy (HRT3) and scanning laser polarimetry (GDxECC) were obtained from each subject. The correlation of average and regional parameters of RNFL and ONH between devices was evaluated.

Results: The RNFL thickness was 90.9 µm when measured with OCT, 24.6 µm with HRT and 48.1 µm with GDx. There was a high correlation between the disc and cup measurements with the HRT and OCT and the RNFL thickness of the OCT and GDx (r > 0.5). A statistically significant correlation was found between RNFL measurements of the HRT and OCT in the superior, temporal and inferior quadrants. Optical coherence tomography (OCT) signal strength correlated with the image quality parameters of the HRT and GDx. The percentage of good quality images was the lowest with the GDx.

Conclusions: The RNFL thickness in Northern Finland birth cohort was at a lower level compared to other studies. The study confirms the difference in measuring ONH parameters between the imaging devices. However, significant correlations between devices were found in the cup volume and cup disc area ratio parameters of the OCT and HRT. The correlations between image quality parameters and glaucoma detection parameters were relatively low.

Kokoelmat
  • Avoin saatavuus [38618]
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen
 

Selaa kokoelmaa

NimekkeetTekijätJulkaisuajatAsiasanatUusimmatSivukartta

Omat tiedot

Kirjaudu sisäänRekisteröidy
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen