Performance of Hemochron ACT-LR and ACT+ Test Cuvettes in Monitoring Low to Moderate Heparin Concentrations: An In Vitro Study
Moilanen, Janne; Pada, Marika; Ohtonen, Pasi; Kaakinen, Timo; Pokela, Matti; Savolainen, Eeva-Riitta; Erkinaro, Tiina (2024-12-04)
Moilanen, Janne
Pada, Marika
Ohtonen, Pasi
Kaakinen, Timo
Pokela, Matti
Savolainen, Eeva-Riitta
Erkinaro, Tiina
Elsevier
04.12.2024
Janne Moilanen, Marika Pada, Pasi Ohtonen, Timo Kaakinen, Matti Pokela, Eeva-Riitta Savolainen, Tiina Erkinaro, Performance of Hemochron ACT-LR and ACT+ Test Cuvettes in Monitoring Low to Moderate Heparin Concentrations: An In Vitro Study, Journal of Cardiothoracic and Vascular Anesthesia, Volume 39, Issue 2, 2025, Pages 447-452, ISSN 1053-0770, https://doi.org/10.1053/j.jvca.2024.12.001
https://creativecommons.org/licenses/by/4.0/
© 2024 The Author(s). Published by Elsevier Inc. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
https://creativecommons.org/licenses/by/4.0/
© 2024 The Author(s). Published by Elsevier Inc. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
https://creativecommons.org/licenses/by/4.0/
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:oulu-202501071059
https://urn.fi/URN:NBN:fi:oulu-202501071059
Tiivistelmä
Abstract
Objectives:
According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.
Design:
An in vitro study.
Setting:
A tertiary care university hospital.
Participants:
Twenty healthy adult volunteers.
Interventions:
Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.
Measurements and Main Results:
All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.
Conclusions:
The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.
Objectives:
According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.
Design:
An in vitro study.
Setting:
A tertiary care university hospital.
Participants:
Twenty healthy adult volunteers.
Interventions:
Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.
Measurements and Main Results:
All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.
Conclusions:
The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.