X-ray scatter in projection radiography
Ylimaula, Satu; Räsänen, Lasse; Hurskainen, Miia; Peuna, Arttu; Julkunen, Petro; Nieminen, Miika Tapio; Hanni, Matti (2023-11-03)
Ylimaula, Satu
Räsänen, Lasse
Hurskainen, Miia
Peuna, Arttu
Julkunen, Petro
Nieminen, Miika Tapio
Hanni, Matti
Oxford University Press
03.11.2023
Satu Ylimaula, Lasse Räsänen, Miia Hurskainen, Arttu Peuna, Petro Julkunen, Miika Tapio Nieminen, Matti Hanni, X-ray scatter in projection radiography, Radiation Protection Dosimetry, Volume 200, Issue 2, February 2024, Pages 120–129, https://doi.org/10.1093/rpd/ncad275
https://creativecommons.org/licenses/by/4.0/
© The Author(s) 2023. Published by Oxford University Press. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
https://creativecommons.org/licenses/by/4.0/
© The Author(s) 2023. Published by Oxford University Press. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
https://creativecommons.org/licenses/by/4.0/
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:oulu-202401041069
https://urn.fi/URN:NBN:fi:oulu-202401041069
Tiivistelmä
Abstract
Projection radiography is the most common radiological modality, and radiation safety of it concerns both radiation workers and the public. We measured and generated a series of scattered radiation maps for projection radiography and estimated effective doses of the supporting person during exposure. Measured adult patient protocols included chest posterior–anterior, chest lateral, pelvis anterior–posterior (AP), abdomen AP and bedside chest AP. Maps concretise spatial distribution and the scattered radiation dose rates in different imaging protocols. Highest and lowest rates were measured in abdomen AP and bedside chest AP protocols, respectively. The effective dose of supporting person in abdomen AP examination at distance of 0.5 m was 300 nSv and in bedside supine chest AP examination at distance of 0.7 m was 0.5 nSv. The estimated annual effective dose of emergency unit radiographer was 0.11 mSv. The obtained effective dose values are small compared to annual dose limits of radiation workers and the public.
Projection radiography is the most common radiological modality, and radiation safety of it concerns both radiation workers and the public. We measured and generated a series of scattered radiation maps for projection radiography and estimated effective doses of the supporting person during exposure. Measured adult patient protocols included chest posterior–anterior, chest lateral, pelvis anterior–posterior (AP), abdomen AP and bedside chest AP. Maps concretise spatial distribution and the scattered radiation dose rates in different imaging protocols. Highest and lowest rates were measured in abdomen AP and bedside chest AP protocols, respectively. The effective dose of supporting person in abdomen AP examination at distance of 0.5 m was 300 nSv and in bedside supine chest AP examination at distance of 0.7 m was 0.5 nSv. The estimated annual effective dose of emergency unit radiographer was 0.11 mSv. The obtained effective dose values are small compared to annual dose limits of radiation workers and the public.