Multipoint Raman spectrometer based on a time-resolved CMOS SPAD sensor

Talala, Tuomo; Kaikkonen, Ville A.; Virta, Eetu; Mäkynen, Anssi J.; Nissinen, Ilkka (2023-04-19)

 
Avaa tiedosto
nbnfi-fe20230929137860.pdf (839.5Kt)
nbnfi-fe20230929137860_meta.xml (37.71Kt)
nbnfi-fe20230929137860_solr.xml (31.88Kt)
Lataukset: 

URL:
https://doi.org/10.1109/LPT.2023.3268333

Talala, Tuomo
Kaikkonen, Ville A.
Virta, Eetu
Mäkynen, Anssi J.
Nissinen, Ilkka
Institute of Electrical and Electronics Engineers
19.04.2023

Talala, T., Kaikkonen, V. A., Virta, E., Mäkynen, A. J., & Nissinen, I. (2023). Multipoint Raman Spectrometer Based on a Time-Resolved CMOS SPAD Sensor. In IEEE Photonics Technology Letters (Vol. 35, Issue 11, pp. 629–632). Institute of Electrical and Electronics Engineers (IEEE). https://doi.org/10.1109/lpt.2023.3268333

https://rightsstatements.org/vocab/InC/1.0/
© 2023 The Author(s). This work is licensed under a Creative Commons Attribution 4.0 License. For more information, see https://creativecommons.org/licenses/by/4.0/.
https://rightsstatements.org/vocab/InC/1.0/
doi:https://doi.org/10.1109/LPT.2023.3268333
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe20230929137860
Tiivistelmä

Abstract

The potential of in-line Raman spectrometers for process monitoring applications has been shown for many industrial processes, but in most cases, only one measurement point has been monitored by one spectrometer. In this letter, we describe and demonstrate a novel, time-resolved method for measuring Raman spectra and fluorescence lifetimes from multiple points using a single excitation source and a single spectrometer. This technique is based on a combination of a time-resolved CMOS SPAD (single-photon avalanche diode) line sensor and a fitting optical light guiding system. The line sensor is designed to make multiple individual measurements at intervals of tens of nanoseconds and the optical light guiding system, in turn, produces matching temporal differences for optical signals from different measurement points. Thereby, signals from different points are distinguished in the time domain. A combined Raman and fluorescence lifetime monitoring of two measurement points was demonstrated with an oil-ethanol emulsion sample.

Kokoelmat
  • Avoin saatavuus [37532]
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen
 

Selaa kokoelmaa

NimekkeetTekijätJulkaisuajatAsiasanatUusimmatSivukartta

Omat tiedot

Kirjaudu sisäänRekisteröidy
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen