The effect of out-of-plane patient shielding on CT radiation exposure and tube current modulations : a phantom study across three vendors

Rautiainen, Jari; Juntunen, Mikael A.K.; Kotiaho, Antti O. (2022-03-21)

 
Avaa tiedosto
nbnfi-fe2023022328370.pdf (1.408Mt)
nbnfi-fe2023022328370_meta.xml (33.13Kt)
nbnfi-fe2023022328370_solr.xml (35.44Kt)
Lataukset: 

URL:
https://doi.org/10.1093/rpd/ncac032

Rautiainen, Jari
Juntunen, Mikael A.K.
Kotiaho, Antti O.
Oxford University Press
21.03.2022

Jari Rautiainen, Mikael A K Juntunen, Antti O Kotiaho, THE EFFECT OF OUT-OF-PLANE PATIENT SHIELDING ON CT RADIATION EXPOSURE AND TUBE CURRENT MODULATIONS: A PHANTOM STUDY ACROSS THREE VENDORS, Radiation Protection Dosimetry, Volume 198, Issue 4, March 2022, Pages 229–237, https://doi.org/10.1093/rpd/ncac032

https://rightsstatements.org/vocab/InC/1.0/
© The Author(s) 2022. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com. This is a pre-copyedited, author-produced version of an article accepted for publication in Radiation Protection Dosimetry following peer review. The version of record Jari Rautiainen, Mikael A K Juntunen, Antti O Kotiaho, THE EFFECT OF OUT-OF-PLANE PATIENT SHIELDING ON CT RADIATION EXPOSURE AND TUBE CURRENT MODULATIONS: A PHANTOM STUDY ACROSS THREE VENDORS, Radiation Protection Dosimetry, Volume 198, Issue 4, March 2022, Pages 229–237 is available online at: https://doi.org/10.1093/rpd/ncac032
https://rightsstatements.org/vocab/InC/1.0/
doi:https://doi.org/10.1093/rpd/ncac032
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe2023022328370
Tiivistelmä

Abstract

The aim of this study was to evaluate how out-of-plane patient shielding affects radiation exposure parameters and tube current modulation on different vendors’ computed tomography (CT) scanners. Helical CT scans were performed using two homogenous phantoms to mimic patient attenuation. Four CT scanners from three vendors were investigated by varying the distance of the patient shield from the border of the imaging volume. Scans were performed with a shield placed before and after the localizer. Changes in volume computed tomography dose index (CTDIvol), dose-length product (DLP) and tube current-time products were studied. Out-of-field lead shield increased the CTDIvol and DLP values for each scanner at least for one scan setting when the shield was present in the localizer. The most notable changes were recorded with >1.3 pitch values when the shield was closest to the scanned volume (2.5 cm), and the scan direction was towards the shield. The usage of patient shields in the localizer CT scans can disturb TCM even when placed 7.5 cm away from the edge of the scan.

Kokoelmat
  • Avoin saatavuus [38618]
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen
 

Selaa kokoelmaa

NimekkeetTekijätJulkaisuajatAsiasanatUusimmatSivukartta

Omat tiedot

Kirjaudu sisäänRekisteröidy
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen