Harmonization of technical image quality in computed tomography : comparison between different reconstruction algorithms and kernels from six scanners

Juntunen, Mikael A.K.; Rautiainen, Jari; Hänninen, Nina E.; Kotiaho, Antti O. (2022-04-05)

 
Avaa tiedosto
nbnfi-fe2022090257115.pdf (9.141Mt)
nbnfi-fe2022090257115_meta.xml (35.5Kt)
nbnfi-fe2022090257115_solr.xml (40.24Kt)
Lataukset: 

URL:
https://doi.org/10.1088/2057-1976/ac605b

Juntunen, Mikael A.K.
Rautiainen, Jari
Hänninen, Nina E.
Kotiaho, Antti O.
IOP Publishing
05.04.2022

Juntunen, M. A. K., Rautiainen, J., Hänninen, N. E., & Kotiaho, A. O. (2022). Harmonization of technical image quality in computed tomography: Comparison between different reconstruction algorithms and kernels from six scanners. Biomedical Physics & Engineering Express, 8(3), 037002. https://doi.org/10.1088/2057-1976/ac605b

https://rightsstatements.org/vocab/InC/1.0/
This is a peer-reviewed, un-copyedited version of an article accepted for publication/published in Biomedical Physics & Engineering Express. IOP Publishing Ltd is not responsible for any errors or omissions in this version of the manuscript or any version derived from it. The Version of Record is available online at 10.1088/2057-1976/ac605b.
https://rightsstatements.org/vocab/InC/1.0/
doi:https://doi.org/10.1088/2057-1976/ac605b
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe2022090257115
Tiivistelmä

Abstract

Purpose: The radiology department faces a large number of reconstruction algorithms and kernels during their computed tomography (CT) optimization process. These reconstruction methods are proprietary and ensuring consistent image quality between scanners is becoming increasingly difficult. This study contributes to solving this challenge in CT image quality harmonization by modifying and evaluating a reconstruction algorithm and kernel matching scheme.

Methods: The Catphan 600 phantom was scanned with six different CT scanners from four vendors. The phantom was scanned with volumetric CT dose indices (CTDIvols) of 10 mGy and 40 mGy, and the data were reconstructed using 1 mm and 5 mm slices with each combination of reconstruction algorithm, body region kernel, and iterative and deep learning reconstruction strength. A matching scheme developed in previous research, which utilizes the noise power spectrum (NPS) and modulation transfer function (MTF), was modified based on our organization’s needs and used to identify the matching reconstruction algorithms and kernels between different scanners.

Results: The matching paradigm produced good matching results, and the mean ± standard deviation (median) matching function values for the different acquisition settings were (a value of 1 indicates a perfect match): CTDIvol 10 mGy, 1 mm slice: 0.78 ± 0.31 (0.94); CTDIvol 10 mGy, 5 mm slice: 0.75 ± 0.33 (0.93); CTDIvol 40 mGy, 1 mm slice: 0.81 ± 0.28 (0.95); CTDIvol 40 mGy, 5 mm slice: 0.75 ± 0.33 (0.93). In general, soft reconstruction kernels, i.e., noise-reducing kernels that reduce sharpness, of one vendor were matched with the soft kernels of another vendor, and vice versa for sharper kernels.

Conclusions: Combined quantitative assessment of NPS and MTF allows effective strategy for harmonization of technical image quality between different CT scanners. A software was also shared to support CT image quality harmonization in other institutions.

Kokoelmat
  • Avoin saatavuus [40265]
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen
 

Selaa kokoelmaa

NimekkeetTekijätJulkaisuajatAsiasanatUusimmatSivukartta

Omat tiedot

Kirjaudu sisäänRekisteröidy
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen