Effect of different ECG leads on estimated R–R intervals and heart rate variability parameters

Jeyhani, Vala; Mäntysalo, Matti; Noponen, Kai; Seppänen, Tapio; Vehkaoja, Antti (2019-10-07)

 
Avaa tiedosto
nbnfi-fe202003248944.pdf (493.6Kt)
nbnfi-fe202003248944_meta.xml (37.98Kt)
nbnfi-fe202003248944_solr.xml (38.37Kt)
Lataukset: 

URL:
https://doi.org/10.1109/EMBC.2019.8857954

Jeyhani, Vala
Mäntysalo, Matti
Noponen, Kai
Seppänen, Tapio
Vehkaoja, Antti
Institute of Electrical and Electronics Engineers
07.10.2019

V. Jeyhani, M. Mäntysalo, K. Noponen, T. Seppänen and A. Vehkaoja, "Effect of Different ECG Leads on Estimated R–R Intervals and Heart Rate Variability Parameters," 2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany, 2019, pp. 3786-3790. doi: 10.1109/EMBC.2019.8857954

https://rightsstatements.org/vocab/InC/1.0/
© 2019 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
https://rightsstatements.org/vocab/InC/1.0/
doi:https://doi.org/10.1109/EMBC.2019.8857954
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe202003248944
Tiivistelmä

Abstract

Heart rate and heart rate variability parameters provide important information on sympathetic and parasympathetic branches of autonomous nervous system. These parameters are usually extracted from electrocardiograms often measured between two electrodes and called an ECG lead. Besides systems intended only for heart rate measurement, ECG measurement devices employ several well-known lead systems including the standard 12-lead system, EASI lead system and Mason-Likar systems. Therefore, the first step is to select the appropriate lead for heart rate variability analysis. The appropriate electrode locations for single-lead measurement systems or the preferred measurement lead in multi-lead measurement are choices that the user needs to make when the heart rate variability is of interest. However, it has not been addressed in the literature, if the lead selection has an effect on the obtained HRV parameters. In this work, we characterized the amount of deviation of heart rate and heart rate variability parameters extracted from nine ECG leads, six from EASI leads and three modified limb leads. The results showed a deviation of 2.04, 2.88, 2.06 and 3.45 ms in SDNN, rMSSD, SD1 and SD2, respectively. A relative difference up to 10% was observed in HRV parameters for single signal frames. Additionally, the discrimination of the R-peaks by amplitudes was evaluated. The A-S lead appeared to have the best performance in all the tests.

Kokoelmat
  • Avoin saatavuus [37647]
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen
 

Selaa kokoelmaa

NimekkeetTekijätJulkaisuajatAsiasanatUusimmatSivukartta

Omat tiedot

Kirjaudu sisäänRekisteröidy
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen