A preliminary study in neonatal cardiorespiratory monitoring through diaphragmatic electromyography

Tobón-Cardona, Marcela; Seppänen, Tapio; Torres, Róbinson (2017-06-13)

 
Avaa tiedosto
nbnfi-fe2019090426586.pdf (259.5Kt)
nbnfi-fe2019090426586_meta.xml (37.97Kt)
nbnfi-fe2019090426586_solr.xml (36.07Kt)
Lataukset: 

URL:
https://doi.org/10.1007/978-981-10-5122-7_273

Tobón-Cardona, Marcela
Seppänen, Tapio
Torres, Róbinson
Springer Nature
13.06.2017

Tobón-Cardona M., Seppänen T., Torres R. (2018) A Preliminary Study in Neonatal Cardiorespiratory Monitoring through Diaphragmatic Electromyography. In: Eskola H., Väisänen O., Viik J., Hyttinen J. (eds) EMBEC & NBC 2017. EMBEC 2017, NBC 2017. IFMBE Proceedings, vol 65. Springer, Singapore

https://rightsstatements.org/vocab/InC/1.0/
© Springer Nature Singapore Pte Ltd. 2018. This is a post-peer-review, pre-copyedit version of an article published in EMBEC & NBC 2017. Joint Conference of the European Medical and Biological Engineering Conference (EMBEC) and the Nordic-Baltic Conference on Biomedical Engineering and Medical Physics (NBC), Tampere, Finland, June 2017. The final authenticated version is available online at: https://doi.org/10.1007/978-981-10-5122-7_273.
https://rightsstatements.org/vocab/InC/1.0/
doi:https://doi.org/10.1007/978-981-10-5122-7_273
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe2019090426586
Tiivistelmä

Abstract

Cardiorespiratory monitoring continues to be a challenge in the Neonatal Intensive Care Unit due to the sensitivity of current techniques to false alarms and artifacts. Surface diaphragmatic electromyography (EMGdi) can be used to measure directly the respiratory muscles activity and to reduce motion artifacts. We propose to acquire ECG and EMGdi with the electrodes placed on the diaphragm to measure heart and respiratory rate in neonates. Data from fourteen neonates was analyzed to compare the measurements of the signal acquired in the diaphragm with the ECG bipolar leads and a reference respiration signal from piezoelectric respiratory effort belt. Bland-Altman plots showed agreement between heart rate monitoring in the diaphragm with no significant differences. Detection of breaths showed similar results in the EMGdi derived wave compared with the reference signal. Thus, heart and respiration rate monitoring in neonates can be accomplished with ECG and EMG signals acquired from the diaphragm.

Kokoelmat
  • Avoin saatavuus [37744]
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen
 

Selaa kokoelmaa

NimekkeetTekijätJulkaisuajatAsiasanatUusimmatSivukartta

Omat tiedot

Kirjaudu sisäänRekisteröidy
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen