The establishment of local diagnostic reference levels in endoscopic retrograde cholangiopancreatography : a practical tool for the optimisation and for quality assurance management

Saukko, E.; Henner, A.; M. T. Nieminen, M. T. Nieminen; Ahonen, S.-M. (2016-03-01)

 
Avaa tiedosto
nbnfi-fe2019121848695.pdf (290.1Kt)
nbnfi-fe2019121848695_meta.xml (32.64Kt)
nbnfi-fe2019121848695_solr.xml (32.34Kt)
Lataukset: 

URL:
http://doi.org/10.1093/rpd/ncw018

Saukko, E.
Henner, A.
M. T. Nieminen, M. T. Nieminen
Ahonen, S.-M.
Oxford University Press
01.03.2016

E. Saukko, A. Henner, M. T. Nieminen, S.-M. Ahonen, THE ESTABLISHMENT OF LOCAL DIAGNOSTIC REFERENCE LEVELS IN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY: A PRACTICAL TOOL FOR THE OPTIMISATION AND FOR QUALITY ASSURANCE MANAGEMENT, Radiation Protection Dosimetry, Volume 173, Issue 4, April 2017, Pages 338–344, https://doi.org/10.1093/rpd/ncw018

https://rightsstatements.org/vocab/InC/1.0/
© The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com
https://rightsstatements.org/vocab/InC/1.0/
doi:https://doi.org/10.1093/rpd/ncw018
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi-fe2019121848695
Tiivistelmä

Abstract

Fluoroscopic procedures are an area of special concern in relation to radiation protection. The aim of this study was to describe the current level of patient radiation doses in endoscopic retrograde cholangiopancreatography (ERCP) collected from a single centre, as well as to establish and review local diagnostic reference levels (DRLs) in ERCP. A total of 100 patients’ radiation doses in ERCP were recorded, and the third-quartile method was adopted to establish local DRLs for ERCP. The mean dose area product (DAP) was 2.05 Gy cm², fluoroscopy time (FT) 1.7 min and the number of images was 3. The proposed local DRLs for ERCP were 3.00 Gy cm² and 3.0 min. Local DRLs were reviewed in a sample of 25 patients 5 y after they had been established. In reviewing data, the averages of DAP and FT were below the local DRLs. Local DRLs help in the optimisation process of fluoroscopic procedures and guides to a good clinical practice.

Kokoelmat
  • Avoin saatavuus [40204]
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen
 

Selaa kokoelmaa

NimekkeetTekijätJulkaisuajatAsiasanatUusimmatSivukartta

Omat tiedot

Kirjaudu sisäänRekisteröidy
oulurepo@oulu.fiOulun yliopiston kirjastoOuluCRISLaturiMuuntaja
SaavutettavuusselosteTietosuojailmoitusYlläpidon kirjautuminen